Pump Magazine On-Line

INDUSTRIAL DIGEST

 

 

 

 

Tested signals electro-magnetic frequencies emitted from the

DETA-AP unit (made by DETA-ELIS company)

Emitted signals were tested and reported on April 3, 2013

by

3505 Francis Circle, Alpharetta, GA 30004

Ph: 770-740-0717

www.ustech-lab.com

Power = 3VDC Battery

No emissions were found below 9kHz. All data presented is collected between 9 kHz and 2 MHz. A noise floor reading was collected before the start of the evaluation. During the transmission the a screen shot was collected. The screen shows the emissions being generated by the unit.

 

The plots below show tested energy levels at various frequencies as emitted from this device. DETA ELIS company claims these DETA-ELIS Programs were focused for certain medical objectives, consistent with Dr. Rife's theory. We have no claims on any effects, benefits, detriments, or otherwise, of these data as carrying any validity and posting these here only for information purpose for the curious scientists or researchers. US Tech company is an official certified laboratory to test various electrical, radio, sound, and similar signals emitted from devices to make sure these are not in violation of the official US standards and guidance's. US Tech was given the DETA-ELIS unit strictly to determine if anything is emitted from it, and if so, at which frequencies. The results of their testing is shown below for DETA-AP unit, and a similar linked page is constructed for the DETA-RITM unit.

 

Note that DETA-ELIS literature claims the DETA-AP unit emits frequencies between 0.1 to 72000 Hz, - which is very different from what the testing of these devices has shown (nothing below 9 kilohertz, and very small signal harmonics above that). This makes the validity of the DETA-ELIS claims questionable.

 

Test results:

 

Vertical axis shows electro-magnetic energy measured in dBm, with the following conversion formula to milliwatts:

milliwatts = 10(dBm/10)

 

i.e. a zero dBm is 1 milliwatt, i.e. a very low energy level. For example, Fig. 1 (horizontal scale = 200 kHz steps; vertical scale = -10 dBm stepping down intervals) shows several peaks at various frequencies, starting at approximately 100 kilohertz, then 200 kilohertz, and so on, with the highest peak shown at 586 kHz and -60 dBm (a small fraction of a milliwatt) power strength. (as a rough guide, such low power is roughly in the range of the energy of a typical radio transmitter).

 

On figures below: "Program" is the number assigned by DETA-ELIS to one of the nearly 3000 program by DETA-ELIS, while "unit number" is one of the 13 programs loaded on the emitter unit for the test.

 

DETA-ELIS Program 01 (P1 unit)= "drainage therapy"

 

DETA-ELIS Program 02 (P2 unit) "basic parasites, 1st day treatment"

 

 

DETA-ELIS Program 03 (P3 unit) "basic parasites, 2nd day treatment"

 

 

DETA-ELIS Program 04 (P4 unit) "parasites of stomach and intestines"

 

DETA-ELIS Program 05 (P5 unit) "parasites detox"

 

DETA-ELIS Program 03 (P6 unit) "regulation of immunize system"

 

 

DETA-ELIS Program 04 (P7 unit) "drainage of liver toxins"

 

 

DETA-ELIS Program 05 (P8 unit) "drainage of kidneys toxins"

 

DETA-ELIS Program 06 (P9 unit) "support of liver and kidneys"

 

DETA-ELIS Program nn (P10 unit) "cleansing of blood"

 

 

DETA-ELIS Program nn (P11 unit) "cleansing of blood plasma"

 

DETA-ELIS Program 13 (P12 unit) "onihomichose"

 

DETA-ELIS Program 04 (P13 unit) "helico-bacteria"

 

DETA-ELIS Program 02 (P14 unit) "drainage of lymphatic system"

 

 

DETA-ELIS Program 13 (P15 unit) "orvi, or3, flu"

 

 

DETA-ELIS Program 19 (P16 unit) "antiseptic"

 

DETA-ELIS Program 20 (P17 unit) "inflammation"

 

DETA-ELIS Program 01 (P18 unit) "detox of complete body"

 

 

It is of interest to have feedback from any researchers familiar with Dr. Rife work and technology behind it. Based on the tested data, what meaning does it have? We will post your comments below.

 

Note that DETA-ELIS literature claims the DETA-RITM unit emits frequencies between 0.1 to 72000 Hz, - which is very different from what the testing of these devices has shown (nothing below 9 kilohertz, and very small signal harmonics above that). This makes the validity of the DETA-ELIS claims questionable.

    

Pumps' Vibration Analysis and Electromagnetic Biomedicine